Part Details | LAMINAR CLEAN AIR WORK STATION
3694-01-377-7575 An assembly designed to produce a purified, positive air flow in a controlled area. This is accomplished by filtering air from outside the station and creating a higher (positive) pressure inside. The Laminar indicates a streamline flow past an object so that the velocity at any fixed point is consistent or varies in a regular manner. Excludes WORK STATION, CLEAN AIR, NON-LAMINAR.
Alternate Parts: 17G4631761, 17G463176-1, 3694-01-377-7575, 01-377-7575, 3694013777575, 013777575
Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
---|---|---|---|
36 | JUL 24, 1993 | 01-377-7575 | 38777 ( WORK STATION, CLEAN AIR, LAMINAR ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 3694-01-377-7575
Part Number | Cage Code | Manufacturer |
---|---|---|
17G463176-1 | 88277 | BOEING COMPANY, THEDBA BOEING |
Technical Data | NSN 3694-01-377-7575
Characteristic | Specifications |
---|---|
GENERAL CHARACTERISTICS ITEM DESCRIPTION | ITEM PROVIDES A CLASS 100/ISOLATED WORKS SURFACE WHICH CAN BE USED WITH A MICROSCOPE. THE DSIGN ISOLATES VIRTUALLY ALL VIBRATION TRANSMISSIONS FROM THE FILTER/BLOWER CONSOLE TO THE SURFACE. IT MAINTAINS A CLASS 100 ENVIRONMENT UTILIZING HEPA FILTRATION. IT HAS A SOLID-STATE SPEED CONTROL FOR ADJUSTING ARFLOW VELOCITY FROM 40 TO 110 FPM. AVE. CFM AT 90 FPM; 900; POWER REQ. 115VAC, 60HZ, 10, 7.5A; FILTRATION, HEPA FILTERS (99.99% EFF TO .3U); DIM, O/A 51"W X 65"H X 44"D; WORKAREA, 49"W X 30"H X 24"D; WEIGHT, 590 LBS |