Part Details | PATIENT THERMOREGULATOR
6530-01-177-2639 An item designed to lower or raise the body temperature of the patient during surgery, as well as quickly reducing and controlling dangerously high temperatures of postoperative infections. Temperature readings can be made esophageally or rectally by the use of thermistor sensing probes.
Alternate Parts: G248, G-248, MTA4700D, MTA-4700D, 6530-01-177-2639, 01-177-2639, 6530011772639, 011772639
Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
---|---|---|---|
65 | APR 21, 1984 | 01-177-2639 | 28787 ( THERMOREGULATOR, PATIENT ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6530-01-177-2639
Part Number | Cage Code | Manufacturer |
---|---|---|
G-248 | 64616 | DEFENSE MEDICAL STANDARDIZATIONBOARD |
MTA-4700D | 30958 | GAYMAR INDUSTRIES INC |
Technical Data | NSN 6530-01-177-2639
Characteristic | Specifications |
---|---|
GENERAL CHARACTERISTICS ITEM DESCRIPTION | AUTO AND MANUALLY OPERATED,110/220V,50/60 HZ,AC,SOLID STATE; 1150 WATTS;IN REUSABLE CONTAINER;SELF-CONTAINED OR ON CART/STAND;MOUNTED ON CASTERS(HEAT/COOL),AND REFRIGERATION SYS;3 MODES;AUTO OPERATION MONITORS PATIENT TEMP AND HEATS OR COOLS PATIENT TO PRESET TEMP;MANUAL OPERATION CONTROLS PRESET FLUID TEMP GOING TO BLANKET;MONITORING OPERATION MONITORS PATIENT TEMP BUT DOES NOT HEAT,COOL,OR CIRCULATE FLUID;25-41 DEG C PATIENT TEMP RANGE;PROVIDED W/VISUAL DISPLAYS TO INDICATE NORMAL OPERATION,VISIBLE/AUDIBLE ALERT INDICATORS TO INDICATE MALFUNCTION,POWER CORD W/PLUG,ESOPHAGEAL AND RECTAL SENSING PROBE ASSEMBLY,14 BLANKETS OF ASSORTED SZ,DRAIN AND TUBE FOR COOLANT/HEATING FLUIDS;8 FT PAD EXTENSION TUBING;FOR USE IN REGULATING PATIENT TEMP BY APPLICATION OF HYPOTHERMIA PADS/BLANKETS THROUGH WHICH TEMP CONTROLLED WATER IS CIRCULATED.DEPLOYABLE MED ITEM ELECTRICAL CHARACTERISTICS,PROVISIONING AND PACKAGING. |
PART NAME ASSIGNED BY CONTROLLING AGENCY | THERMOREGULATOR: PATIENT AUTO&MANUAL 115/220V 50/60 HZ AC |