Part Details | OXYGEN ANALYZER
6630-01-006-6100 An item used for the detection and measurement of the oxygen content in air or other gases.
Alternate Parts: 0A3502, 0A350-2, 6630-01-006-6100, 01-006-6100, 6630010066100, 010066100
Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
---|---|---|---|
66 | JUL 19, 1975 | 01-006-6100 | 27944 ( ANALYZER, OXYGEN ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6630-01-006-6100
Part Number | Cage Code | Manufacturer |
---|---|---|
0A350-2 | 51317 | MELOY LABORATORIES INC |
Technical Data | NSN 6630-01-006-6100
Characteristic | Specifications |
---|---|
DETECTION RANGE IN PERCENT | P0.0/P100.0 |
CARRYING HANDLE | NOT INCLUDED |
CURRENT TYPE | AC |
VOLTAGE IN VOLTS | 115.0 |
FREQUENCY IN HERTZ | B50.0$B60.0 |
PHASE | SINGLE |
ANALYSIS CELL TYPE | RAPID RESPONSE |
CORROSION RESISTANT FEATURE | INCLUDED |
SAMPLE TYPE | FLOWING |
RESPONSE TIME | 15.0 SECONDS |
OVERALL WIDTH | 19.000 INCHES NOMINAL |
OVERALL DEPTH | 20.000 INCHES NOMINAL |
OVERALL HEIGHT | 12.000 INCHES NOMINAL |
SPECIAL FEATURES | RANGES: 0 TO .01 PPM,0 TO .1 PPM,0 TO .5 PPM,0 TO 1 PPM,0 TO 5 PPM,0 TO 10 PPM; MIN DETECTABLE SENSITIVITY .0005 PPM; RELATIVE HUMIDITY RANGE 10 TO 95 PCT.; PRECISION PORM 2 PCT.; OPERATIONAL PERIOD 7 DAYS UNATTENDED; AMBIENT TEMP RANGE 10 TO 40 DEG C; MEASURES CONCENTRATION OF OZONE IN AMBIENT AIR BY PHOTOMETRICALLY MONITORING THE CHEMILUMINESCENCE RESULTING FROM THE REACTION OF ETHYLENE GAS WITH OZONE; BUILT-IN CALIBRATION SYSTEM |