Part Details | LEAK STANDARD,GAS D
6635-01-379-6339
Alternate Parts: 922801HELIUM, 922-801HELIUM, OM8, OM-8, 6635-01-379-6339, 01-379-6339, 6635013796339, 013796339
Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
---|
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6635-01-379-6339
Part Number | Cage Code | Manufacturer |
---|---|---|
922-801HELIUM | 15718 | VACUUM INSTRUMENT CORPORATIONDBA VIC LEAK DETECTION |
OM-8 | 15718 | VACUUM INSTRUMENT CORPORATIONDBA VIC LEAK DETECTION |
Technical Data | NSN 6635-01-379-6339
Characteristic | Specifications |
---|---|
GENERAL CHARACTERISTICS ITEM DESCRIPTION | LEAK RATE RANGE (ATM-CC/SEC): 1 TO 9 X 10-8; HELIUM GAS; SUPPLIED WITH 1/2 IN. OD SMOOTH TUBE OUTLET FITTING;OUTSIDE MATL 304 STAINLESS STEEL CYLINDER: OVERALL DIM 14.5 IN. LG; 2.0 IN. IN. DIA; 20 LBS; ORIFICE TYPE LEAK; USES PRECISION TAPERED GLASS CAPILLARY AS LEAK ELEMENT; PROVIDES GRADUAL PRESSURE RISE AT PREDICTABLE RATE; INDIVIDUALLY CALIBRATED USING ACCUMULATION OVER TIME METHOD IAW AMERICAN VACUUM SOCIETY STANDARD 2.2-1968 |
FUNCTIONAL DESCRIPTION | CALIBRATES HELIUM LEAK DETECTOR, P/N 1120HE/1038-17 PRIOR TO TESTING FIRE EXTINGUISHER |